7.6. Menu 4 Filter & Motion Detection
4
Digital filter & No Motion setup (Menu 4.1 to 4.4)
4.1
Low pass filter cut off frequency (command FL)
-
Settings: 0 - 8 with UP/DOWN buttons
4.1.x
Cut off frequency:
IIR mode
FIR mode
4.1.0
No digital filter
No digital filter
4.1.1
18 Hz
19.7 Hz
4.1.2
8 Hz
9.8 Hz
4.1.3
4 Hz
6.5 Hz
4.1.4
3 Hz
4.9 Hz
4.1.5
2 Hz
3.9 Hz
4.1.6
1 Hz
3.2 Hz
4.1.7
0.5 Hz
2.8 Hz
4.1.8
0.25 Hz
2.5 Hz
4.2
Digital filter Mode - IIR or FIR (command FM)
-
Choose IIR or FIR
4.3
Update rate and averaging (command UR)
4.3.x
Averaging (from 1 to 128 readings)
4.3.0
0 - each reading
4.3.1
1 - average of 2 readings
4.3.2
2 - average of 4 readings
4.3.3
3 - average of 8 readings
4.3.4
4 - average of 16 readings
4.3.5
5 - average of 32 readings
4.3.6
6 - average of 64 readings
4.3.7
7 - average of 128 readings
4.4
Motion detection
4.4.1
No motion range (value range from 1 to 65 535 d) (command NR)
Weight value changes within this range will be considered as
'stable'
4.4.2
No motion time (value range from 1 to 65 535 ms) (command NT)
Time span for the no motion detection where the signal has to be
'stable'