7.7. Menu 5 – Analog Output
Analog output setup (Menu 5.1 to 5.5)
Weight value for minimum analog output (command AL)
Set the weight value which corresponds to minimum output
Examples for scale 0 … 3 000kg
Minimum 0kg or with 600kg preload
0kg = 4mA - setting 00000
600kg = 4mA - setting 00600
0kg = 0mA - setting 00000
600kg = 0mA - setting 00600
Weight value for maximum analog output (command AH)
Set the weight value which corresponds to maximum output
Examples for scale 0 … 3 000kg
3 000kg = 20mA - setting 03000
3 000kg = 20mA - setting 03000
Analog output base (command AA)
gros - analogue output follows Gross value
net - analogue output follows Net value
PEA - analogue output follows Peak value (Maximum)
AUEr - analogue output follows Average value
HoLd - analogue output follows Hold value
PP - analogue output follows Peak - Peak value
UALL - analogue output follows Valley value (Minimum)
dISP - analogue output follows Display value
oFF - analogue output is switched OFF
Analog output mode (command AM)
Test current output range 0…20mA
Set a mA value as a test output signal, independent of the
applied load/weight - e.g. 004.000 for 4mA